Can CORE tell me my core temperature in the heat?
Not to a threshold: in 19 elite hockey players, CORE read 0.06-0.34 °C below an ingested pill on average, 41-60% of readings were >0.3 °C off, and the gap widened as core rose.
Goods et al. (Murdoch University, Western Australian Institute of Sport, the Australian Institute of Sport, Hockey Australia and the University of Western Australia) asked whether the CORE wearable sensor, which estimates core temperature from the skin, agrees with an ingested telemetric pill when core temperature is high.
Nineteen players of the Australian national women's field hockey squad wore CORE and took a BodyCAP pill through 4 training sessions in hot conditions at a heat camp before the 2020 Olympic Games. Lin's concordance was "poor" in every session.
CORE read lower than the pill in all sessions, by 0.06-0.34 °C on average, with what the abstract calls wide mean 95% confidence intervals of ±0.35 to ±0.56 °C. A smoothed regression showed the error was not constant: CORE under-read more as core temperature rose. In every session 41-60% of all samples differed by more than ±0.3 °C.
The authors conclude that CORE is not a valid alternative to a telemetric pill, particularly during exercise in the heat when core temperature is expected to be high.
Our reading: in elite athletes training in the heat, a CORE reading was within ±0.3 °C of the pill about half the time or less, and it erred low exactly where a heat-training target or a safety limit sits; from the abstract, so the sensor version and the conditions are not known.
Protocol
Key Results
| Outcome | Result (abstract) |
|---|---|
| Concordance (Lin's CCC) | "Poor" agreement in all 4 sessions |
| Mean bias, CORE minus pill | −0.06 °C to −0.34 °C (CORE lower in every session) |
| Width around the bias | "Mean 95% confidence intervals" (the abstract's term) ±0.35 °C to ±0.56 °C |
| Error vs core temperature | Non-linear; greater under-reading as core temperature rose (LOESS) |
| Samples beyond ±0.3 °C | 41-60% of all samples in each session |
Mechanisms
Practical Application
Limitations
How this entry was checked and changed
from the abstract; a validation study with n = 19 elite athletes and a valid criterion, which the rubric places at B; held at B- because only the abstract could be read (the open-access full text is behind a bot check), so the sensor version, firmware, placement, conditions, sample counts and the funding and conflict statements are unknown