Accuracy of Respiratory Gas Variables, Substrate, and Energy Use from 15 CPET Systems During Simulated and Human Exercise
Van Hooren et al. tested the accuracy of 15 widely used CPET systems by connecting each to a metabolic simulator that mimics breath-by-breath gas exchange; they then checked the simulation results with people cycling and, for some of the systems, measured reliability between sessions.
Systems (manufacturer in brackets): Quark CPET and K5 (COSMED); MetaLyzer 3B and MetaMax 3B (Cortex); Vyntus CPX and Oxycon Pro (Vyaire); Omnical (Maastricht Instruments); Ergocard Clinical, Ergocard Pro and Ultima (MGC Diagnostics); PowerCube Ergo (Ganshorn/Schiller); Ergostik (Geratherm); VO2masterPro (VO2master); PNOĒ; Calibre (Calibre Biometrics).
Protocol
Simulated breath-by-breath gas exchange; each system's minute ventilation, breathing frequency, VO2, VCO2, RER, energy from carbohydrate and fat, and total energy expenditure compared to the simulated values.
Key Results — absolute percentage errors during simulation (range across systems)
Conclusion (abstract): error differed substantially between systems, with only a few systems showing a consistently acceptable error. Which systems those were is not named in the abstract.