Precision and accuracy of four handheld blood lactate analyzers across low to high exercise intensities
Against a Biosen reference between 0.88 and 4.89 mM, the four handheld meters (two units each of Lactate Plus, Lactate Pro2, Lactate Scout 4 and TaiDoc TD-4289) read on average −7% (Plus), 7% (Pro), −10% (Scout) and 42% (Tai), with 95% prediction intervals 0.60-0.87 mM wide at 3 mM; precision was poor for all four, and even the second stationary analyzer (YSI) read −32%: lactate values from different meters are not comparable
The Norwegian Olympic Sports Center compared measured blood lactate concentrations from four commonly used handheld lactate analyzers with two stationary analyzers, on venous samples taken at exercise intensities from low to high.
All four handheld meters had poor precision compared with the stationary analyzers; their mean offsets from the Biosen differed by brand and direction, and the most precise handheld meter (TaiDoc) was the least accurate. Content from the abstract only.
Participants
Not described in the abstract (venous blood samples taken at exercise intensities ranging from low to high). The unit of comparison is the paired measurement, not the athlete.
Protocol
Key Results
| Analyzer | Mean relative difference to Biosen | Residual standard error vs Biosen | 95% prediction interval width at 3 mM (Biosen) |
|---|---|---|---|
| Lactate Plus | −7% | 0.18 mM | 0.72 mM |
| Lactate Pro2 ("Pro") | 7% | 0.20 mM | 0.80 mM |
| Lactate Scout 4 | −10% | 0.22 mM | 0.87 mM |
| TaiDoc TD-4289 ("Tai") | 42% | 0.15 mM | 0.60 mM |
| YSI Sport 1500 (stationary) | −32% | 0.06 mM | 0.23 mM |