Can my Tymewear find my VT1 and VT2?
Roughly at VT1, low at VT2: in 15 recreationally active adults the Tyme Wear shirt put VT2 0.8-1.3 km/h and 5-8 bpm below gas analysis; limits of agreement reached 35 bpm. Strap, cycling untested.
Gouw et al. asked whether the Tyme Wear smart shirt, which estimates minute ventilation from chest-wall stretch and uses it to locate the first and second ventilatory thresholds, is reliable and valid against laboratory gas exchange.
Nineteen recreationally active adults did two treadmill ramp tests 7-10 days apart, following the maker's app-guided protocol (walk 3 min at 3.0 mph, then run from 3.3 mph, +0.3 mph every 30 s to exhaustion), wearing the shirt while a Parvo Medics TrueOne 2400 measured gas exchange; two exercise physiologists located VT1 and VT2 by eye from the ventilatory equivalents.
Fifteen were analysed. The shirt's thresholds repeated well from test to test (ICC 0.91-0.98 for time and speed, 0.78-0.84 for heart rate), similar to gas exchange (0.90-0.96).
Against gas exchange the shirt put VT1 on average 0.2-0.3 mph slower and about 4 bpm lower (significant only for heart rate in trial 1), and VT2 0.5-0.8 mph slower and 5-8 bpm lower (significant in both tests), with wide individual limits of agreement (VT2 speed up to −1.5 to +3.0 mph, VT2 heart rate −18.7 to +35.1 bpm). The four fittest participants were under-read most.
The authors conclude the shirt is reliable but less valid than the lab, adequate for setting zones in most recreational users, with caution for fitter people.
Our reading: a small, one-generation test and the only check of a Tymewear threshold in this base; VT1 from the shirt sat near the lab value on average, VT2 was systematically low and individually scattered, and nothing here speaks to the strap, to cycling or to ventilation itself.
Protocol
Key Results
Agreement, gas exchange (TRUE) minus shirt (S-PRED); positive = shirt lower (Bland-Altman, Results 3.4; km/h our conversion at 1 mph = 1.609 km/h):
| Threshold | Measure | Trial 1: bias ± SD (limits of agreement) | Trial 2: bias ± SD (limits of agreement) |
|---|---|---|---|
| VT1 | Time point | 0.3 ± 0.9 min (−1.4 to 2.0) | 0.4 ± 1.4 min (−2.2 to 3.1) |
| VT1 | Treadmill speed | 0.2 ± 0.5 mph (−0.8 to 1.2); ≈ 0.3 km/h (−1.3 to 1.9) | 0.3 ± 0.8 mph (−1.3 to 1.8); ≈ 0.5 km/h (−2.1 to 2.9) |
| VT1 | Heart rate | 4.0 ± 7.2 bpm (−10.1 to 18.1) | 4.3 ± 11.1 bpm (−17.6 to 26.1) |
| VT2 | Time point | 0.82 ± 1.46 min (−2.0 to 3.7) | 1.2 ± 1.8 min (−2.4 to 4.8) |
| VT2 | Treadmill speed | 0.5 ± 0.9 mph (−1.2 to 2.2); ≈ 0.8 km/h (−1.9 to 3.5) | 0.8 ± 1.1 mph (−1.5 to 3.0); ≈ 1.3 km/h (−2.4 to 4.8) |
| VT2 | Heart rate | 5.3 ± 8.6 bpm (−11.7 to 22.2) | 8.2 ± 13.7 bpm (−18.7 to 35.1) |
Means (Table 4), TRUE vs shirt: VT1 speed 4.6 vs 4.4 mph in both trials, heart rate 130.8 vs 126.8 bpm (trial 1, significant) and 129.7 vs 125.4 (trial 2); VT2 speed 7.1 vs 6.6 and 7.1 vs 6.4 mph, heart rate 163.5 vs 158.2 and 162.3 vs 154.0 bpm, time 10.5 vs 9.1 and 10.5 vs 9.3 min (all VT2 differences significant).
Mechanisms
Practical Application
Limitations
How this entry was checked and changed
single-lab validation study, n = 15 analysed of 19 tested, read from the full text (open access, CC BY 4.0). The rubric places a validation study at B; held at B- because 4 of 19 participants were dropped for "equipment error or data loss" without saying which device failed (a shirt that fails in some people and is then excluded looks better than it is), and parts of the reliability and discussion numbers do not add up (see Limitations)