Optical (PPG) pulse-rate variability differed from ECG heart-rate variability at rest by a fraction of the ECG standard deviation on average
In a meta-analysis of 43 validation studies in healthy people, 8 of which could be pooled per measure, the gap between study means was 0.188 of the ECG standard deviation for RMSSD (95% CI 0.066 to 0.309) and 0.134 for SDNN (0.014 to 0.255, PPG reading higher); only resting or controlled recordings were pooled, no limits of agreement could be computed, no accepted threshold says how small an error is small enough, and agreement looked more variable in sleep, exercise and stress
Xu and colleagues asked how far heart-rate variability measured from an optical pulse signal (PPG, as in watches, rings, arm bands, finger clips and phone cameras) departs from the same variability measured from an ECG, in healthy non-clinical people.
They searched three databases from 2017 to May 2026 and kept 43 studies that reported RMSSD or SDNN against an ECG reference; only 10 gave the paired means and SDs needed for pooling, and one representative condition per study was chosen by a fixed hierarchy that preferred resting, seated or supine, short-term recordings.
Because Bland–Altman limits of agreement were not reported consistently, the pooled outcome was the absolute difference between the PPG and ECG group means divided by the ECG SD. For RMSSD this was 0.188 SD (95% CI 0.066 to 0.309; I² 11.69%) over 8 studies and for SDNN 0.134 SD (0.014 to 0.255; I² 0%) over 8 studies.
With direction kept, PPG read higher than ECG for SDNN (0.134, p = 0.029) and showed a non-significant upward trend for RMSSD (0.110, 95% CI −0.054 to 0.275). The 33 studies that could not be pooled suggested better agreement in controlled resting or paced-breathing recordings and more variable agreement in sleep, exercise and recovery, stress tasks and field recordings.
The authors conclude that PPG pulse-rate variability may serve for within-device trend monitoring under controlled conditions, that the pooled figures do not extend to sleep, exercise, stress or daily life, and that it is not interchangeable with ECG HRV.
Participants
Protocol
Key Results
| Analysis (PPG vs ECG, resting or controlled) | Studies | Estimate (units of ECG SD) | 95% CI | I² |
|---|---|---|---|---|
| RMSSD, absolute standardized error | 8 | 0.188 | 0.066 to 0.309 | 11.69% |
| SDNN, absolute standardized error | 8 | 0.134 | 0.014 to 0.255 | 0% |
| RMSSD, signed (positive = PPG higher) | 8 | 0.110, p = 0.188 (not significant) | −0.054 to 0.275 | 51.14% |
| SDNN, signed | 8 | 0.134, p = 0.029 | 0.014 to 0.255 | 0% |
| RMSSD, criterion ECG only | 6 | 0.178 | 0.029 to 0.327 | 22.9% |
| RMSSD, without the paced-breathing study | 7 | 0.192 | 0.048 to 0.335 | 23.85% |
| SDNN, criterion ECG only (chest-strap references removed) | 5 | 0.103 (not significant) | −0.048 to 0.254 | 0% |
| SDNN, without one study at a time | 7 | 0.108 to 0.156; CI crossed 0 after removing either of two studies | — | — |
| RMSSD Pearson correlation (exploratory) | 6 | r = 0.980 | 0.847 to 0.998 | 97.33% |
Practical Application
Limitations
How this entry was checked and changed
read in full, but the review applied no risk-of-bias tool and no certainty grading (only a descriptive reporting checklist, 70-85% per study), so the rubric's condition for A- is not met; in addition the pooled metric compares group means and gives no individual limits of agreement, and each pooled analysis rests on 8 small studies (HRV is the measure under validation, so no marker step)