Is my smart scale's body-fat number accurate?
Not for one reading: vs a lab 4-compartment model, the best consumer scales could miss by about ±6.5 points, most foot-to-foot ones by ±11-15, a 3-month change by ±3.5-5 (our arithmetic).
Siedler et al. asked how repeatable consumer bioimpedance scales are, how close their body-fat percentage comes to a laboratory four-compartment model, and whether they track change.
Seventy-three healthy adults, standardised for fasting, hydration and exercise, stood on 15 devices in random order — one hand-to-hand device, ten foot-to-foot scales and four octapolar (hands and feet) models, three of them consumer-grade and one research-grade — and were measured by a 4C model built from air-displacement plethysmography (body volume), DXA (bone mineral) and bioimpedance spectroscopy (total body water).
Thirty-seven came back 12-16 weeks later. Immediate test-retest precision was high (precision error 0.0-0.49%), but in five foot-to-foot scales suspiciously so (repeat readings never more than 0.1% apart), which the authors read as the device repeating its last number for the same user.
Against the 4C model the constant error ran from −3.5 to +11.7 body-fat points; only five devices were within ±2 points of it at the group level (the research-grade Seca mBCA, InBody H20N, Withings Body Cardio, Tanita BC-554 and Tanita UM-081).
For change over 12-16 weeks, when the true mean change was +0.2 points (SD 2.9), the devices' constant error was small (−0.4 to +1.3) but individual agreement was poor (standard error of the estimate 1.7-2.6 points, concordance 0.37-0.78), and seven devices compressed change systematically.
Our reading: a well-standardised study with a gold-standard criterion; for an athlete, a single body-fat reading from a scale carries an individual error of several points even on the best models, and a three-month change of a few points is inside the noise of every device tested.
Protocol
Key Results
Body-fat percentage, each device vs the 4C model (Tables 2 and 3; constant error = device minus 4C, in body-fat points):
| Device (configuration) | Cross-sectional constant error | Total error | Equivalent ±2 | Limits of agreement, single reading (our arithmetic) | Change over 12-16 wk: constant error | Total error of change | Limits of agreement for change (our arithmetic) |
|---|---|---|---|---|---|---|---|
| Seca mBCA 515/514 (octapolar, research) | −0.2 | 3.3 | Yes | −6.7 to +6.3 | 0.3 | 2.1 | −3.8 to +4.4 |
| InBody H20N (octapolar) | −0.1 | 3.3 | Yes | −6.6 to +6.4 | −0.4 | 2.1 | −4.5 to +3.7 |
| Tanita BC-568 InnerScan (octapolar) | −1.4 | 4.4 | No | −9.6 to +6.8 | 0.7 | 1.9 | −2.8 to +4.2 |
| Omron HBF-516 (octapolar) | 3.3 | 5.6 | No | −5.6 to +12.2 | 0.8 | 1.9 | −2.6 to +4.2 |
| Withings Body Cardio (foot-to-foot) | −0.4 | 5.8 | Yes | −11.8 to +11.0 | 0.6 | 1.9 | −3.0 to +4.2 |
| Tanita BC-554 Ironman (foot-to-foot) | 0.7 | 5.8 | Yes | −10.7 to +12.1 | 0.6 | 2.0 | −3.2 to +4.4 |
| Tanita UM-081 (foot-to-foot) | 0.6 | 5.8 | Yes | −10.8 to +12.0 | 0.7 | 1.9 | −2.8 to +4.2 |
| Omron HBF-306 (hand-to-hand) | −3.5 | 5.4 | No | −11.6 to +4.6 | 0.2 | 2.1 | −4.0 to +4.4 |
| Other foot-to-foot scales (RENPHO, HAWANA, Wyze, INEVIFIT, Weight Watchers, Vitagoods) | −1.4 to +3.6 | 6.8-7.8 | No | about ±12-15 around the constant error | −0.3 to +0.3 | 2.1-2.6 | about ±4-5 |
| Seca 804 (foot-to-foot) | 11.7 | 12.6 | No | +2.5 to +20.9 | 1.3 | 2.9 | −4.0 to +6.6 |
The paper prints each device's limits of agreement only inside its Bland-Altman figures.
The limits above are constant error ± 1.96 × the SD of the differences, with that SD recovered from the total error (and cross-checked against the 95% CI of the constant error); this reproduces the cross-sectional SDs the abstract gives (4.1 for the Omron HBF-306, 4.7 for the Seca 804) and its change SD of 2.1 for the InBody H20N; for the Seca 804's change the reconstruction gives 2.6, so the abstract's 2.7 is used.
Our arithmetic, not the authors' figures.
Mechanisms
Practical Application
Limitations
How this entry was checked and changed
cross-sectional measurement study with a gold-standard criterion, n = 73 (37 re-tested after 12-16 weeks), read from the full text (open access, CC BY 4.0). The rubric places a cross-sectional or test-validation study at B and needs ≥100 for B+; one lab, healthy non-athletes, and the devices' "athlete" settings were not used